AV500 – SHIM UPDATE
Some users had complained that they are not able to get good line shape for their sample and Topshim software did not help, apparently, either. My experience today has been different. The spectrum shown below is the result of a successful shimming operation with Topshim, for running this reasonably demanding 1D spectrum that suppresses 95% […]
Posted in Spectrometer, Usage | No Comments
DRX499 – UPDATE ON LINESHAPE
Users had noticed that the lineshape on this instrument has been rather poor. This is primarily due to the fact that the magnet is still taking its own time to settle down after the cryoshimming and B0 field adjustment approximately a month ago. Supercon Magnets are notoriously unpredictable when it comes to their expected duration […]
Posted in Uncategorized | No Comments
GENERAL ALERT ! COMPRESSED AIR INTERRUPTION TOMORROW, Wed, 25th Jun
Facilities and services has announced a 2 hour window of compressed air disruption from 5.00 am to 7.00 am on Wedensday, 25th June ’14. The walk-on users won’t be affected by this event but those who want to plan overnight runs should take note that the VT system will be non-functional during this event. Please […]
Posted in Uncategorized | No Comments
UPDATE ! DRX499 – SYSTEM BACK TO NORMAL
Please let us know if you face any issues again.
Posted in Hardware, Spectrometer | No Comments
ALERT – DRX499 – COMMUNICATION SOFTWARE BEING REINSTALLED
Normal acquisition is not proceeding due to a corruption of low level software on the host computer. This is being re-installed and the system should be up soon.
Posted in Software, Spectrometer | No Comments