Spectrometer

AV800: system down due to gradient malfunction

We are waiting for a hardware replacement part.  The expected timeline for the same is end of this week i.e. 8 Feb 2013.  Any changes to this schedule will be posted here.

Posted in Hardware, Spectrometer | No Comments »

AV700: UPDATE

A pneumatic control board failed in the AV700 console, making the spectrometer unusable.  We have been waiting for a replacement of the said board from the manufacturer. The current status is that, this board should arrive at the facility in roughly two weeks time.  Once the system is up and functioning, I will post another […]

Posted in Hardware, Spectrometer | No Comments »

AV301 : UPDATE ON VT

The chiller unit is malfunctioning and therefore variable temperature is not available on the instrument until further updates.  All spectra need to be recorded at 298 K i.e. room temperature.

Posted in Hardware, Spectrometer | No Comments »

AV301 : PROBE UPDATE – 19F probe back

The erstwhile damaged 19F probe has been repaired in house and has been tested to function satisfactorily.  We thank the users for their patience and understanding while waiting for the 19F probe to be available.  Please let me or Paul know if you observe anything of significance with the TXD 1H {19F,13C} probe.

Posted in Hardware, Spectrometer | No Comments »

AV301 UPDATE : NEW AUTOMATED BROADBAND PROBE

The 19F probe has been compromised and will need extensive repairs. In place of the old probe, the new BBI probe is installed in the system. IMPORTANT NOTE : This is a new generation probehead that does NOT allow you to do manual tune/match i.e. WOBB. You must use the command ‘atma’  instead and the […]

Posted in Hardware, Software, Spectrometer | No Comments »

AV500 UPDATE : SYSTEM IS BACK TO NORMAL

Users can start reserving time at once and use the instrument. Thank you for your patience.

Posted in Host Computer, Operating System, Software, Spectrometer | No Comments »

AV301 : ALERT – 19F CHANNEL MALFUNCTION

A probe head related issue is being looked into .  If it cannot be fixed within the next hour, an alternative will be provided. Please wait for further updates.

Posted in Hardware, Spectrometer | No Comments »

CAUTION : DISRUPTION CAUSED BY USER REBOOT

Users are strongly reminded that,  it is best to avoid the temptation of rebooting the computer,  when they ‘think’ that that will remedy a given situation.   The requirement to reboot a computer to take care of any issue is quite rare and any problems that are faced by users require much simpler measures such as […]

Posted in Host Computer, Operating System, Usage | No Comments »

AV800 : REMINDER – WALK ON TIME NEXT WEEK

Please contact me by before the end of the day tomorrow i.e. Friday, 2nd Nov., to avail the remaining time during the Walk-On period in the 800 MHz spectrometer.

Posted in Spectrometer, Usage | No Comments »

AV500 UPDATE : BF1 SHIFT COMPLETE

The Base Frequency of the spectrometer is now reset to : 499.956572 MHz. Getting the line shape close to factory spec is still a work in progress. But, the current resolution achieved is comparable to the values with which we were operating prior to the frequency shift.  Existing user created parameter sets and the Bruker […]

Posted in Software, Spectrometer, Usage | No Comments »