UPDATE : AV700 USERS – SYSTEM READY FOR USE
The BBO-1H{X}-Z gradient probe is in place now. Please Remember: After creating a dataset, do ‘edsp’ to re-initialize the new base frequency. Load the default shim file : stdbbo and then start the shimming process. NMR Facility Manager 6/4/2018
Posted in Spectrometer, Usage | No Comments »
UPDATE ! AV700 – SYSTEM BACK IN OPERATION – NEW BASE FREQUENCY ! PLEASE READ
The Lock system is back to normal. PLEASE NOTE THE FOLLOWING IMPORTANT CHANGE ! New System Base Frequency : 699.962448 MHz. This reconfiguration is needed to compensate for the long term loss of energy by the Supercon Magnet. How does this affect you ? If you simply open your existing data set and start copying […]
Posted in Software, Spectrometer, Usage | No Comments »
UPDATE ! AV700 USERS – INSTRUMENT IS OPERATIONAL WITH SOLUTIONS NMR PROBEHEAD
The current probehead is TXI-1H {13C,15} with Z-gradient. The nominal VT range is -60C to +70C without any special care or considerations. The probehead is capable of reaching to a larger VT limits but there will be an accompanying reduction in the continuous NMR observation time to protect the magnet and shim electronics. Please contact […]
Posted in Spectrometer, Usage | No Comments »
AV700 SOLIDS NMR USAGE : BRIEF CLARIFICATION
Unlike the solutions spectrometers, solid state NMR instrument is not amenable to a walk-on mode of operation, by the very nature of this spectroscopy. First of all, the technique requires the use of an expensive rotor (approx. cost $1000/- ) in place of the NMR tube that costs about $10/-. The sample preparation also needs […]
Posted in NMR Applications, Spectrometer, Usage | No Comments »
ALERT ! AV700 USERS – IMMINENT CHANGE – SOLUTIONS TO SOLIDS STARTING NEXT WEEK
The 700 will switch to solid state NMR mode beginning April 16th, Monday. It is expected that the instrument will remain in solids operation for a minimum of 2 weeks and it might extend to a month. All solutions users should take note and plan accordingly. NMR Facility Manager 4/11/2018
Posted in Spectrometer, Usage | No Comments »
AV700 – J-YOUNG SAMPLE TUBE USERS – PLEASE READ !
I am picking up from my last post on the Sample Handling change in AV700. If you bring a J-Young tube, please confirm its straightness with the special Aluminum tester that is atop the NMR console from now on, we recommend you to use only the Ceramic Spinner, which balances the center of mass closer […]
Posted in Spectrometer, Usage | No Comments »
IMPORTANT NOTE ! AV700 USERS – SAMPLE HANDLING, PLEASE READ
To tackle the recurrent issue of J-Young sample tubes getting stuck inside the magnet, we are coming up with an alternative that shows promise in addressing the issue. J-Young Tubes : You must only use the ceramic spinner shown below from now on : Regular NMR Tubes : You must use the PEEK PLASTIC spinner […]
Posted in Hardware, Spectrometer, Usage | Comments Off on IMPORTANT NOTE ! AV700 USERS – SAMPLE HANDLING, PLEASE READ